Operator manual
Corival
22
changed, in the example mentioned above: S. Pressing the
▼▲
keys give you the ability to
toggle between the possibilities for programming a stage in a protocol: S, T, L, H, P and
Recovery.
S= step
the workload can be ad or – and the duration of this hyperbolic
stage can be determined; the load is independent of the RPM.
T= torque
fixed torque: the torque in Nm can be adjusted with certain duration.
L= linear
the workload is coupled with RPM with constant
α
as the adjustable
parameter for the desired duration; load P =
α
* N
2
[W] goes up with
square of cadence.
H= heart rate
the workload is regulated as a function of the adjustable heart rate for a
certain duration in this heart rate controlled step.
P= proportional
the workload is increased linear as a function of time (ramp protocol). It
will take the total duration time to reach the programmed workload;
R = Repeat
repeat steps (until max. workload): you have to enter the workload step and
the duration of the step only once.
Recovery
the recovery stage is the last stage of a protocol with a fixed workload and
an unlimited duration (only 1 recovery stage in 1 protocol)
After the confirmation of the workload during the recovery stage the menu PROGRAM BPM
will appear. Toggle with the
▼▲
keys for the measurement possibilities:
No BPM
no BPM module available, select this option when the BPM is not installed, or
no automatic blood pressure measurement during this protocol;
Repeat
allows you to program a repeat time in minutes during the exercise stages:
EOS
End Of Stage; automatic measurement at every end of the exercise step with
the possibility to program the measurement time before the end of the stage.
Recovery time
If you have programmed automatic BP measurement in the protocol, you get
the possibility to program the repeat time of the BP measurement during the recovery stage:
After the programming of the BPM the menu PROTOCOL will appear again. You can RUN
EDIT or DELETE one of the protocols now.
B.3.7.2
Example Programming Protocol
For the purpose of clarification we will program a demonstration protocol with all
possibilities. Execute the following steps to get into the programming mode.
1.
Select PROTOCOL in the MAIN MENU, confirm
( )
2.
Select PROTOCOL edit, confirm
( )
3.
Select New, confirm
( )
4.
Give a protocol name: test
5.
With the
▼▲
keys you can scroll through the alphabet confirm, use next
( )
to go to
the next position
6.
The program screen appears, the underscore showing the programming parameter: