Operator manual
Corival
12
3.4.2
Keyboard On/Off
If you do not want to interrupt a training or you do not want the test subject to interrupt the
training by pressing a key accidentally, the possibility is offered to disable the keyboard
Now you wonder how to get back the control to the keyboard. Act as follows:
Switch off the ergometer
Switch it on again while keeping the
( )
button pressed. Now you have back the control. Go to
SYSTEM PARAMETERS / KEYBOARD ON/OFF and set it to ON.
At the next startup of the ergometer the keyboard is enabled again.
3.4.3
Test Functions
For servicing and calibration of the device a number of diagnostic tools is available in the
software. They are hidden under Test functions. You need a safety code to enter the
TESTFUNCTIONS.
3.4.4
P In/Out Analog
The ergometer can be controlled by an external device which has an analog output only. This
is done using the ANALOG mode. With P In/out Analog the analog scaling can be adjusted:
- 1 V analog in - can correspond with a workload of 100 W / 200 W / 500 W.
3.4.5
TC Bytes
The Test code (TC) Bytes offer in case of malfunction or poor performance of the BP option
the possibility to make error codes visible. Your local distributor can help you diagnose the
problem with the Error Code.
3.4.6
RS232 Protocol
When the ergometer is externally controlled via RS232 a protocol should be selected here.
3.4.7
Language
You can choose different languages for the ergometer.
3.4.8
Device No
When RS232 communication is used with the ergometer, it is identified with a so called
device number. This offers the possibility of looping through ergometers.