29
Malfunction
Possible Cause
Corrective
No display appears on the
screen
Main power cable is
disconnected
Check that the main power
cable is connected to an AC
source.
Main switch is not turned on
Turn on the main switch,
located on the base of the
Front panel.
Fuse break
Check the fuse in power
inlet.
Pressure not displayed on
User Interface
Pneumatic supply not
connected
Check pneumatic supply and
connections
Reservoir inner cable is
disconnected
Check inner cable
connection from Reservoir
EQ RC50 Integrated
Dispenser I-4.0 is defective
Contact Henkel Service
No start signal
XS1 cable connection is not
secure.
Switch the power switch to
the position O (OFF).
Tighten the screws of the
plug. Switch the power
switch to the position I (ON).
Footswitch defective
Replace the Footswitch
No communication.
Check the networking or call
Henkel Service.
No product, too little or too
much product
Dispensing pressure not set
correctly.
Adjust dispensing pressure
setting.
Pressure hose not properly
connected.
Connect air pressure hose
correctly.
Luer-Lock tip cap not
removed.
Replace Luer-Lock tip cap
with a dispensing needle.
Dispensing needle clogged,
too small or too large.
Replace dispensing needle.
Dispensing valve not
correctly connected or
defective.
Check the dispensing valve
(see instruction manual for
dispensing valve).