Compared the difference between the actual value of the digital regulator and
preset ‘5 bar’, then click “+”/ “
-
“(
7)
to increase/decrease the value until the
reading of “
actual value of digital regulator
”
is
“5 bar”.
-
Click the “
”
(5)
to record the actual reading as
2nd “reference” point.
-
Press button Save to save all adjusted parameters.
Note: The display reading is in bar and prop. regulator is in mPa (factor of 10, e.g., 5
bar = 0.5 mPa). If a difference in the calibration values is displayed, repeat the
calibration procedure to do calibration again.
6
Application Hints
As with all adhesives, performance depends on conditions of use. Suggestions or
recommendations contained herein are for guidance only since actual conditions of
use are outside the supplier’s
control.
6.1 Shutdown for Longer Periods of Non-use (>recommended idle time)
•
Disconnect the pneumatic supply from the controller.
•
Clean the product hose and dispensing valve.
Recommended maximum idle times for different products are shown below:
Adhesive
Maximum idle time for dispensing Systems
Anaerobic
2 weeks
Cyanoacrylate
1 week
UV-acrylate
2 weeks
Acrylate
1 week
Epoxy
2 weeks
Activator
n.a.
Primer
n.a.