5.6 Set-up Configurations
5.6.1 Auto Timer Mode (Single dispense valve)
•
Connect electrical and pneumatic supplies and dispense valve as described in
section 4.2
•
Place adhesive bottle in the reservoir as described in section 4.3
•
Enter Setting page, select
“
”
mode for channel A or Channel B.
•
Click the value
, this can be performed using the direct
numerical input function to set dispensing time for channel A or channel B.
Click the ‘+’/ ‘
-
’ to increase/reduce the val
ue by one step, the dispensing time
is adjustable from 0.0 to 999.9s.
•
Set the reservoir pressure and acceptable range, the pressure can be changed
from 0.00 to 8.00 bar. Optionally set the pressure unit bar/psi if necessary.
•
Set the number of cycles for Low Level Counter that are permitted after low
level is detected. The system will not dispense once low level is detected
when the low level counter function is disabled.
•
Click “
” button to p
ressurise
the reservoir and select “
” run
working mode. The system is ready for dispensing
in “Time” mode
.
5.6.2 Auto Timer Mode (Two dispense valves)
•
Connect electrical and pneumatic supplies and dispense valve as described in
section 4.2
•
Place adhesive bottle in the reservoir as described in section 4.3
•
Enter Setting page, select “
”
run mode for channel A and channel B.
•
Click the value
, this can be performed using the direct
numerical input function to set dispensing time for channel A and channel B.
Click the ‘+’/ ‘
-
’ increase/reduce the value by one step, the dispensing time is
adjustable from 0.0 to 999.9s.
•
Set the reservoir pressure and acceptable range, the pressure can be changed
from 0.00 to 8.00 bar. Optionally set the pressure unit bar/psi if necessary.