SCP System • Replacing components
34
CP_SEM_60-00-00.002
Flow sensor board (DIGIFLOW)
Removal of the
flow sensor
board is necessary in the following cases:
◗
If there is no flow display despite the fact that the flow sensor is correctly connected and intact.
There is no special error message (error code).
◗
For replacement of the mains filter and the connecting cable to the pump control panel.
Note that the Transonic FPT1004 flow sensor board has become obsolete. If the flow sensor is
defective, installed Transonic FPT1004 flow sensor boards must be replaced with the DIGIFLOW flow
sensor board (p/n 90-900-520).
Fig. 17: Dismantling the flow sensor board
◗
Remove the two connectors of the flow sensor
cable from the flow sensor connections.
◗
Unscrew the four screws indicated (Allen key size 2.5) including the washers.
◗
Carefully pull the board upwards.
Then replace the board in the opposite order.