SCP System • Replacing components
CP_SEM_60-00-00.002
35
Connecting cable and/or mains filter
Before removing the connecting cable, the
flow sensor
board must first be dismantled (page 34).
Fig. 18: Dismantling the connecting cable (1)
◗
Unscrew the two marked screws on the mains filter
(Allen key size 2.5) including the washers.
◗
Lift up the mains filter.
◗
Disconnect the connections on the connecting cable from the mains filter (brown cable) and from
the toggle switch (white cable). If you only want to replace the mains filter, remove all connectors
now and ignore the following steps for replacing the cable.
◗
Detach any cable clips that prevent the connecting cable from being removed.
Fig. 19: Dismantling the connecting cable (2)
◗
Open the terminal screws at the CON1 connector for the two protected connecting cables (black,
triple wire and twin wire).
◗
Pull the five cable ends out of the plug (black/yellow/green and yellow/green)