SCP System • Replacing components
CP_SEM_60-00-00.002
37
Dismantling the shaft angle encoder (SAE)
Fig. 22: Dismantling the shaft angle encoder
◗
Pull the rotary knob
upwards.
◗
Release the mounting nut
with an Allen key (size 1.5) and remove the mounting nut.
◗
Release the guide sleeve
by means of a suitable hinged pin wrench. Hold the shaft angle
encoder assembly securely while removing the guide sleeve to prevent damage, e.g. by an
unintentional movement of the shaft angle encoder body.
◗
Unplug the connector
of the shaft angle encoder.
◗
The shaft angle encoder
may now be removed.
Mod. Item
Description
Number
SP number
A
43
Rotary knob
96-190-057
A
44
Mounting nut
28-91-76
A
45
Washer PA 6.2 x 12.0 x 0.5 (see fig. 23)
71-120-514
A
46
Guide sleeve
60-02-02
A
47
Shaft angle encoder
97-103-603
Threaded rod, M3x3
71-118-301
A
48
Washer (see fig. 23)
60-02-04
A
49
Mounting adapter
45-21-45
Содержание 60-00-00
Страница 1: ...SCP System Service Manual ...
Страница 12: ...SCP System System description 12 CP_SEM_60 00 00 002 3 1 2 Overview Fig 1 Overview A B C D ...
Страница 24: ...SCP System System description 24 CP_SEM_60 00 00 002 Page intentionally left blank ...
Страница 28: ...SCP System Error messages and error diagnosis 28 CP_SEM_60 00 00 002 Page intentionally left blank ...
Страница 51: ...SCP System Appendix CP_SEM_60 00 00 002 51 Notes ...
Страница 52: ...SCP System CP_SEM_60 00 00 002 ...