SCP System • Error messages and error diagnosis
26
CP_SEM_60-00-00.002
4.1.2
Error codes
The following error messages are only shown on the 7-segment display “Actual value” (RPM).
Error
Description/Possible causes
Corrective measures
(please observe sequence)
RAM defective
– Replace
CPU
board
EPROM defective, checksum incorrect
– Replace EPROM or
– Replace
CPU
board
EEPROM defective
– Replace
CPU
board
Processor error
– Replace
CPU
board
Display error, either
Display
board or
CPU
board defective
– First replace
display
board; if the error
display persists
– Replace
CPU
board
Processor error
– Replace
CPU
board
Difference set value-actual value of pump
more than 250 RPM,
drive unit error
– Replace
motor control
board
A/D converter: reference voltage outside
tolerance limits
– Replace
CPU
board
Shut-off path test (on powering up)
defective
– Replace
CPU
board
Difference set value-actual value of pump
more than 1000 RPM, drive unit or
CPU
board error
– Replace
motor control
board; if the
error display persists
– Replace
CPU
board
E01
E02
E04
E06
E08
E09
E62
E64
E71
E76
Содержание 60-00-00
Страница 1: ...SCP System Service Manual ...
Страница 12: ...SCP System System description 12 CP_SEM_60 00 00 002 3 1 2 Overview Fig 1 Overview A B C D ...
Страница 24: ...SCP System System description 24 CP_SEM_60 00 00 002 Page intentionally left blank ...
Страница 28: ...SCP System Error messages and error diagnosis 28 CP_SEM_60 00 00 002 Page intentionally left blank ...
Страница 51: ...SCP System Appendix CP_SEM_60 00 00 002 51 Notes ...
Страница 52: ...SCP System CP_SEM_60 00 00 002 ...