SCP System • Error messages and error diagnosis
CP_SEM_60-00-00.002
27
Error
Description/Possible causes
Corrective measures
(please observe sequence)
Pump continues running:
Inconsistent actual value impulses
– Test Hall sensor/
motor control
board
connection (CON 3), clean where
necessary
– If the error display persists (i.e.
defective Hall sensor), replace drive
unit completely
Pump stops:
No actual value impulses
– Test Hall sensor/
motor control
board
connection (CON 3), re-establish where
necessary
– Test
motor control
board/pump control
panel connection (CON 2), test
connecting cable/short circuit, where
necessary replace connecting cable
– Replace
motor control
board
– Replace
CPU
board
see page 42
see page 40
see page 32
Shaft angle encoder defective
– Replace shaft angle encoder
(complete)
– If the error display persists, replace
CPU
board
see page 37
see page 32
Logical program run error
– Replace EPROM
– If the error display persists, replace
CPU
board
see page 32
Pump coding invalid
– Test CAN/24V connection to HLM (e.g.
test with a roller pump)
– Test
CAN/24V
connection in pump
control panel (CON 1), test connecting
cable/short circuit, where necessary
replace connecting cable
– Adjust DIP switch (1) on the circuit
board
CPU
(ZKR 9910) to the correct
setting:
Switch setting ON: when operating
with the S5 System
Switch setting OFF: when operating
with the S3 System
see page 35
see page 29
Data transmission
drive unit/pump control panel defective
– Test
motor control
board/pump control
panel connection (CON 2), test
connecting cable/short circuit, where
necessary replace connecting cable. If
the error display persists:
– Replace
motor control
board
– Replace
CPU
board
see page 35
see page 40
see page 32
E78
E80
E81
E82
E87
Содержание 60-00-00
Страница 1: ...SCP System Service Manual ...
Страница 12: ...SCP System System description 12 CP_SEM_60 00 00 002 3 1 2 Overview Fig 1 Overview A B C D ...
Страница 24: ...SCP System System description 24 CP_SEM_60 00 00 002 Page intentionally left blank ...
Страница 28: ...SCP System Error messages and error diagnosis 28 CP_SEM_60 00 00 002 Page intentionally left blank ...
Страница 51: ...SCP System Appendix CP_SEM_60 00 00 002 51 Notes ...
Страница 52: ...SCP System CP_SEM_60 00 00 002 ...