SCP System • Replacing components
CP_SEM_60-00-00.002
31
5.1.2
Replacing pump control panel components
EPROM
Fig. 14: Replacing the EPROM
contains the operating software (firmware) of the SCP System. Replacement is
necessary in the following cases:
◗
To perform an update
◗
In the event of error messages that indicate a defect in the EPROM (E02, E81).
Observe the following instructions when replacing the EPROM:
◗
Always use suitable removal and insertion tools so as not to bend the contacts.
◗
Avoid static charging (and hence damaging the electronic components) by suitable earthing
measures.
◗
When inserting the new EPROM, ensure the correct alignment (notch pointing towards centre of
board).