SCP System • Replacing components
36
CP_SEM_60-00-00.002
Fig. 20: Dismantling the connecting cable (3)
◗
Open the strain relief clamp on the rear of the casing and pull the cable out.
◗
Remove the two sheaths from the old cable. You will need these to re-attach the new cable.
When inserting the new cable (in the reverse order) ensure that the connections at the CON1 connector
are correctly assigned:
Fig. 21: CON1 connection configuration
1 2 3 4
5 6 7
1
White
2
Brown
Current supply from mains filter.
Caution
: Do not transpose!
3
Green
4
Yellow
from connecting cable (twin wire)
5
Green
6
Yellow
7
Black (sheath)
from connecting cable (triple wire)