www.lionprecision.com/two-year-warranty/
Document No. M017-7500.004
28
Oakdale, MN, USA www.lionprecision.com 651-484-6544
Warm-Up
POSITIONING CAPABILITY
Related Standards:
ISO 230-3, 6
▪
ASME B5.54, B5.57, 7.6.2.1, 7.7.2.1
Process:
1. Start with a ‘cold’ spindle (minimum 12 hours of no operation before beginning test).
2. Select a duration (10-120 minutes)
3. Start the spindle turning at 75% of maximum.
4. Begin the test.
The measurement will not run if the spindle is not rotating.
Description:
Measure the position of the target pin in all three axes. The first reading is set as the zero/
reference for the rest of the test. After the initial reading, position readings are taken every
minute and plotted on the chart. Multiple samples per revolution for 32 revolutions are averaged
to find the static location of the spindle.