27
FDCS/CC Crucible Carrier and
Quartz Crucible
By using the FDCS/CC sample carrier (1), sample
can be loaded into the stage without removing the
stage lid. The crucible carrier is loaded through
the side door.
The crucible carrier has two spring clips to hold
the crucible flat against the silver block. This is
to ensure good thermal contact.
To load the quartz crucible into the carrier, lift the
two spring clips about 2mm upward and slide the
crucible in to the holder.
Release the two spring clips to hold the Crucible
in place
Load the sample into the quartz crucible. Use as
little sample as possible to ensure a small thermal
load and therefore better temperature sensitivity.
Microscope objectives require a flat surface to
give maximum field of view. Place a 13mm
coverslip on top of your sample and gently tap the
top with the back of the vacuum tweezers to dis-
perse powder samples or ensure good thermal
contact of irregularly shaped samples.
1