Connecting the Pirani Gauge
The surfaces where the O
-
ring (1) sits should be
completely free of dirt. The O
-
rings should be
smeared with a small amount of vacuum grease
before loading in between the Pirani Gauge (2)
and the Stage Vacuum Connector (3).
Ensure that the O
-
ring is correctly seated in the
gauge and vacuum connector before tightening
the Clamping Bracket (4) screw (5).
Connect the Pirani Cable with the LAN type
connector (5) to the Pirani and the other end of
the cable to the back of the T95; Slot 4; labelled
‘VAC’, see page 9.
1
2
3
4
5
5