Sample Preparation
and Loading
The stage chamber must first be prepared for
vacuum. Ensure that the chamber is completely
clean and dry inside. Finger grease and dirt must
be cleaned off with IPA (isopropanol).
Sample Loading Using G16.3 Sample
Holder
A Freeze Drying Sample is prepared using a
G16.3 Sample Holder (1).
To load the G16.3 into the FDCS stage, open the
door of the stage by unscrewing the door thumb-
screw. The G16.3 slides into the brass housing
located just inside the door.
The Sample Holder Ramp (2) will guide the
G16.3 up on to the silver block.
Close the door and tighten the thumbscrew to seal
the stage.
Adjust the XY manipulators so the holder is in the
middle of the silver block