22
Using the FDCS196 System
Programming Temperature
Please refer to Linksys32 Software manual to
program the temperature by software using a PC.
Using Linksys32 Software to Control
MV196 Motorised Valve
When the Pirani gauge and / or MV196 System
are connected correctly to the T95, you will see
an additional Vacuum Toolbar in the Linksys32
software main window. See opposite picture.
To set vacuum units click on the Pa icon (1) to
select Pascal’s, or use the pull down list to select
other units.
Set Vacuum
The Pirani gauge will display a live vacuum read-
ing shown in the Main Vacuum Window (2).
To set a vacuum, turn on the Vacuum Pump and
type in a vacuum value (value dependent on unit
chosen) in the Set point’s box (3). Click the
Close Valve icon (4) and the motorised valve will
automatically adjust itself to achieve the set vacu-
um value.
Note: a graphical indicator of the valve position
between the Close and Open position can be seen
in Motorised Vacuum Valve Indicator Bar (5).
Note: to take the FDCS196 back to atmospheric
pressure, turn off the vacuum pump and click on
Open Vale icon (6).
Note: if you have just the standard FDCS196
system with no MV196 system. The Open Valve
and Close Valve icons are greyed out and will not
function.
4
3
5
6
2
1