Vacuum Parameters
Touch the active area (1) to change the main
screen to the Vacuum menu screen.
In the Vacuum main screen there will be 3 options
to control the Vacuum value.
1.
Auto: in this mode the motorised valve will
automatically adjust itself to your programmed
vacuum value.
Touch the
Auto
button (2) and make sure the
word ‘auto’ is displayed under the word ‘Vac’ (3).
Use the keypad to type in a vacuum value and
touch
Enter
button (4) to go back to the main
screen.
Note: The MV196 is only designed to hold a
controlled vacuum between
100mbar to 0.1mbar.
Any values outside this range may not be accu-
rate.
Note: touch the
Cancel
button to go back to the
main screen without changing the vacuum value.
Note: Current:
-
is the current programmed rate.
Note: Range:
-
is the maximum and minimum
vacuum value that can be program for the selected
vacuum unit (mBar or kPa).
2. Atmos: in this mode the motorised valve will
fully open to set the vacuum pump to atmos
-
pheric pressure.
Touch the
Atmos
button (5) and make sure the
word ‘atmos’ (6) is displayed under the word
‘Vac’ to set the Ramp to
Atmos
mode. Touch
Enter
button (7) to go back to the main screen.
1
3
2
4
5
6
7