Unpacking, Inspecting and Installing
248083408-1 Rev 1
3-3
Operation
Understanding the LEDs
The front-panel LEDs have special meanings when illuminated. The
figure below shows location of the LEDs and gives a brief description
of their meaning.
Figure 3-1 Front Panel LEDs
•
The green POWER LED turns on when the instrument is powered
on.
•
The green FLOW LED turns on when the flow is correct.
•
The green FLOW LED will turn off if flow is insufficient.
•
The orange SERVICE LED will stay on steady if Laser power is
low, sensor optics are dirty or the view chamber contains foreign
objects.
•
The blue SAMPLING LED turns on when the instrument is
counting.
Features
The REMOTE 2014P, 3014P or 5014P instrument has the following
features:
1. Optional Remote Display connection (color Touchscreen interface)
2. Four optional 4-20mA output connections
3. DIP switch control for device addressing
Содержание Remote 2014P
Страница 1: ......
Страница 2: ......
Страница 6: ......
Страница 10: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual t iv 248083408 1 Rev 1...
Страница 12: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual ii 248083408 1 Rev 1...
Страница 14: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual 1 2 248083408 1 Rev 1...
Страница 20: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual 2 6 248083408 1 Rev 1...
Страница 40: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual 3 20 248083408 1 Rev 1...
Страница 72: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual B 2 248083408 1 Rev 1...
Страница 74: ...Lighthouse REMOTE 2014P 3014P 5014P Operating Manual C 2 248083408 1 Rev 1...
Страница 79: ......
Страница 80: ......