Page 72
Electrode Connectivity
The electrode connection message appears when there are problems
with the contact of one or more electrodes.
Please check the electrode connections when you see these screens.
When the electrodes are reconnected, it will take up to 3 minutes to
update the screen (assuming that the Bluetooth link is up).
Electrode connectivity message screen examples:
Electrode connectivity message
Screen after pressing OK.
The screen will return to normal
operation after problem is solved.
Please follow these instructions to ensure proper contact of the
electrodes:
1) Make sure that the sensor lead wires are properly connected to the
electrodes.
2) Make sure the electrodes are properly attached.
Please refer to “Electrode Placement” section for further information.
Note
- There are physiological conditions that may affect
the cell phone monitor from detecting the electrode
connection and can last for several hours. Even though
the electrodes seem to be in good contact with the body,
the message will continue to appear. Some examples
are very dry skin or right after attaching the electrodes. If
the condition does not resolve, contact LifeWatch.