Caution
If the sensor (ACT I firmware version 0.1g
and ACT III firmware version 1.0.4) is not
connected to the cell phone monitor (via
Bluetooth) for more than 2 hours (ACT I) or 6
hours (ACT III) and the sensor's battery is
replaced, the sensor (on reconnection) will
start with new data and not download the
data stored, thus overriding the data stored
when it was not connected.
Electrode disconnection might cause a faulty
ECG analysis and/or false events due to
noise created by the electrode
disconnection.
The impedance test (occurring every two
minutes) overrides ECG recording, which
means the ECG will lack 1 second (ACT I) or
0.5 second (ACT III) of recording every two
minutes.
Do not turn the cell phone monitor sound off
or reduce the volume so that it is inaudible.
After exiting the ACT (monitor) application, it
will take up to 3 minutes for all the processes
to end. This means you must wait 3 minutes
before starting the ACT (monitor) application
again.
Take the charged cell phone monitor with
you and wear the sensor at all times (except
when showering or bathing) during the
monitoring period.