Page 41
ACT III Electrode Placement
Attach the white lead wire with electrode under the second right
rib. (3 fingers below the collarbone) towards the breastbone
(center).
Place the red lead wire with electrode in a direct line from the
end of the breastbone that intersects with the imaginary line that
extends down from the midpoint of the collarbone, as shown in
ACT III Electrode Placement figure.
Place the black lead wire with electrode at the point where a
horizontal line from the red electrode intersects with the middle
armpit line.
Note:
For the red and black lead wires, if necessary lift the left breast to
place the electrode properly.
Place the green lead wire with electrode on the left side of the
abdomen.
Figure 8.
ACT III Electrode Placement
Содержание ACT I
Страница 74: ...Page 74 Electrode Messages for ACT III White electrode problem Black electrode problem Red electrode problem...
Страница 101: ...Page 101 THIS PAGE LEFT INTENTIONALLY BLANK...
Страница 102: ...SUP362 ACT Patient User Guide Rev C CR 11 023 PLEASE CALL 1 800 517 6330 FOR 24 7 CUSTOMER SUPPORT...