Starting/Stopping the ACT Sensor
The insertion of the battery starts the sensor and connection to the cell
phone monitor via Bluetooth.
Removing the battery will stop the sensor and end the connection to the
cell phone monitor.
The ECG recording will start only after connection is established with the
cell phone monitor.
No Connection with Cell Phone Monitor
When the sensor disconnects from the cell phone monitor (Bluetooth
disconnection), the sensor will continue to record and store the ECG data
(2 hours for ACT I; 6 hours for ACT III).
In case the disconnection period is longer than the maximum recording
time, the sensor will store the LAST time period (2 hours for ACT I; 6
hours for ACT III) of the disconnection period.
Sensor Sound Prompts
The sensor will signal its status (through beeps), i.e., normal
operation, battery level, system errors, manual events.
Sensor Sound
Description:
1 beep, one-time
Sensor activated
1 long beep then short beep, every
ten minutes
Low battery level warning
1 long beep then short beep,
repetitive
Critical battery level error,
sensor stops recording
3 beeps, repetitive
Critical error, system stops
Continuous sensor sound
Manual event buttons pressed