8. Troubleshooting
Please follow the steps listed below in order to resolve the problem.
If this does not solve the problem contact LifeWatch.
Problem
Possible Cause(s)
Solution
No beep when
the sensor is
activated (up to
10 seconds after
battery
inserted).
Battery not placed
properly/reversed
polarity.
The battery is very
low or depleted.
1) Check the battery
placement and
polarity.
2) Sensor battery needs
to be changed, refer to
“Sensor Battery
Insertion/Replacement
” section, for
information on how to
replace the battery.
Displays an
electrode
problem
(message).
Electrode connection
does not provide
proper contact
(loose, disconnected
or bad contact with
skin).
1) Make sure that the
sensor lead wire is
properly connected to
the electrodes and
sensor (for attachable
leads).
2) Make sure the
electrodes are
properly attached to
the skin.
3) There are
physiological
conditions that may
affect the cell phone
monitor from detecting
the electrode
connection and can
last for several hours.