StoreX
Series
Remote Operation V0712 .doc
26
RKi, 26.11.2020
DISCONTINUED
2.1.2
Monitor Flags
P0
0000
Rot.EN
0500
Pn
0001
GateEN
0501
GtClse
0002
GateEN
0502
GtOpen
0003
Rot.Dir
0503
0004
GtTmOut
0504
F.Door
0005
AccsLED
0505
SW 2E0
0006
SW 2E1
0007
SW 2E2
0008
SW 2E3
0009
Acs.
1100
Key Valid
InPos
Ready 1915
RotPos. DM01
Accs. DM00
AccV. DM02
SetSpd. DM91
aSlpe. DM92
bSlpe. DM93
intvl. DM94
_
Status Flag of the Handling Controller. The status message “0“ means inactive, the status
message “1“ means active.
Code
Status Flag
Comment
P0
0000
0-initiator of carrousel
Pn
0001
Position-initiator of carrousel
GtClse
0002
Gate closed switch
GtOpen
0003
Gate opened switch
KeySw
0004
Key switch
Gn LED
0005
Green LED
SW 2E0
0006
Manual positioning switch bit0
SW 2E1
0007
Manual positioning switch bit1
SW 2E2
0008
Manual positioning switch bit2
SW 2E3
0009
Manual positioning switch bit3
Acs.
1100
Carrousel access (remote or manual)
Key Valid
Manual positioning switch valid
Ready
1915
Ready-Bit
Rot.Pos.
DM1
Actual carrousel position
Access.
DM0
Carrousel set position
AccV.
DM2
Access accepted