StoreX
Series
Remote Operation V0712 .doc
17
RKi, 26.11.2020
DISCONTINUED
t = actual value (Word=16Bit)
s = set value (Word=16Bit)
Note !
Z-height data memory values may be different on 3
rd
generation
instruments due to different reduction ration of the lift drive
Divide DM-values by four for 3
rd
generation instruments
Always read DM values prior any change
1.3.4
Special Commands STP1000
Custom commands are commands defined for a specific application for. These commands
will work on instruments supporting the specific application. These instruments may differ in
their hardware configuration .
Special Commands for STP1000 support addressing the Tube Puncher Module (TPM).
Command
Response
Select external TPM device for plate transfer ST 1104
cr
OK
cr lf
Select standard transfer station for plate
transfer
RS 1104
cr
OK
cr lf
Select lower level at TPM for plate transfer
(Daughter Plate)
ST 1105
cr
OK
cr lf
Select upper level on TPM for plate transfer
(Mother Plate)
RS 1105
cr
OK
cr lf
Read lift transfer height for plate transfer at
TPM
RD DM41
cr
ddddd
cr lf
Read transfer height for plate transfer at
upper level (Mother Plate)
RD DM42
cr
ddddd
cr lf
Read transfer height for plate transfer at
lower level (Daughter Plate)
RD DM43
cr
ddddd
cr lf
Set lift transfer height for plate transfer at
TPM to h1
WR DM41
sp
h1
cr
OK
cr lf
Set plate transfer height for plate transfer at
upper level (Mother Plate) in TPM to h2
WR DM42
sp
h2
cr
OK
cr lf
Set plate transfer height for plate transfer at
lower level (Daughter Plate) in TPM to h3
WR DM43
sp
h3
cr
OK
cr lf
1.3.5
Cassette Configuration Look-up Table
1
Up to 20 types of cassette may be stored in the Cassette Type Look-up Table (CTT). The
Data Memory area of the Cassette Type Look-up Table is DM230..DM250. At this time 14
cassettes types are pre-configured:
Used
TypeNr.
Type
Pitch
Address Content
+
0
MTP
23mm
DM230 788
1
FW Version 7.1 and later only