StoreX
Series
Remote Operation V0712 .doc
10
RKi, 26.11.2020
DISCONTINUED
Get Plate
WR DM0 x
WR DM5 y
ST 1907
Put Plate
WR DM0 x
WR DM5 y
ST 1906
BarCode Search
ST 1710
Short Import
WR DM10 p
Short Export
WR DM15 n
1.2.9
System Status
The Handling response on Ready-Polling (“x”) is ‘0’ (handling not ready to accept
commands) or ‘1’ (handling ready, command may be sent). In its idle status the handling sets
the Ready Bit to ‘1’.
Command
Response
Read Ready-Bit
RD
sp
1915
cr
x
cr lf
When polling the Ready Bit after sending a command, wait at least 200ms before requesting
the Ready Status. We recommend to wait 100ms – 200ms between polling sequences.
A sample program will explain the polling sequence in detail. Note that monitoring a flag
change either from ‘0’ to ‘1’ or from ‘1’ to ‘0’ may also be advisable.