StoreX
Series
Remote Operation V0712 .doc
15
RKi, 26.11.2020
DISCONTINUED
Read Gate Mode 1 Flag (default=0), wait
behind Gate for ST 1901
RD
sp
1601
cr
x
cr lf
Set Gate Mode 1 Flag (default =0)
ST
sp
1601
cr
OK
cr lf
Reset Gate Mode 1 Flag (default =0)
RS
sp
1601
cr
OK
cr lf
Read Gate Mode 2 Flag (default=0), close
Gate after timeout
RD
sp
1602
cr
x
cr lf
Set Gate Mode 2 Flag (default =0)
ST
sp
1602
cr
OK
cr lf
Reset Gate Mode 2 Flag (default =0)
RS
sp
1602
cr
OK
cr lf
Read Gate Mode 3 Flag (default=0), never
close Gate automatically
RD
sp
1603
cr
x
cr lf
Set Gate Mode 3 Flag (default =0)
ST
sp
1603
cr
OK
cr lf
Reset Gate Mode 3 Flag (default =0)
RS
sp
1603
cr
OK
cr lf
Read Input- and Output Transfer Station
selection
RD
sp
1607
cr
x
cr lf
Swap Input and Output Transfer Station
ST
sp
1607
cr
OK
cr lf
Swap Transfer Stations to default position
RS
sp
1607
cr
OK
cr lf
Read Handshake Mode flag (default =0)
RD
sp
1610
cr
x
cr lf
Activate Handshake Mode (default =0)
ST
sp
1610
cr
OK
cr lf
De-activate Handshake Mode (default =0)
RS
sp
1610
cr
OK
cr lf
Read Plate Trace selection
RD
sp
1611
cr
x
cr lf
Plate Trace on, stop on plate handling error
ST
sp
1611
cr
OK
cr lf
Plate Trace off, do not check plate
RS
sp
1611
cr
OK
cr lf
Read Wait Plate selection
RD
sp
1612
cr
x
cr lf
Wait Plate on, stop at x-fer station until plate
present resp. x-fer station is free
ST
sp
1612
cr
OK
cr lf
Wait Plate off, do not check x-fer station
status
RS
sp
1612
cr
OK
cr lf
Read Plate Verify selection
RD
sp
1613
cr
x
cr lf
Plate verify on, stop at x-fer station to allow
BCR read
ST
sp
1613
cr
OK
cr lf
Plate verify off
RS
sp
1613
cr
OK
cr lf
Read User Door Lock door
1
status
RD
sp
1701
cr
x
cr lf
Lock User door
2
ST
sp
1701
cr
OK
cr lf
Unlock User door
3
RS
sp
1701
cr
OK
cr lf
Read LED / Beeper Alarm status
RD
sp
1702
cr
x
cr lf
Activate LED- / Beeper Alarm
ST
sp
1702
cr
OK
cr lf
De-activate LED- / Beeper Alarm
RS
sp
1702
cr
OK
cr lf
Do BarCode Search Scan
ST
sp
1710
cr
OK
cr lf
Scan one cassette
4
ST
sp
1711
cr
OK
cr lf
Turn BCR on
ST
sp
1712
cr
OK
cr lf
Turn BCR off
ST
sp
1713
cr
OK
cr lf
BCR Reset
ST
sp
1714
cr
OK
cr lf
Read Actual Slot Position n (0..2)
RD
sp
DM1
cr
nnnnn
cr lf
Read Handler z-Offset (default = 600)
RD DM20
cr
ddddd
cr lf
Read Handler dz Pick- & Place-Movement in
Stacker (default = 500)
RD DM21
cr
ddddd
cr lf
Read Handler In-Transfer z-Position (default
~42’000)
RD DM22
cr
ddddd
cr lf
Read Handler z-Pitch (default =1925)
RD DM23
cr
ddddd
cr lf
1
available if door lock option present
2
available if door lock option present
3
available if door lock option present
4
Requires 2-way communication feature