Troubleshooting
Light Source / Sensor Problems
Using the LI-6400
20-31
2
Blown fuse?
The flow board fuse protects the lamp. If it is blown, a number of things will
not work, including the lamp, lamp fan, pump and cooling fans.
3
Check the lamp voltage
Unscrew the lamp connector (Figure 20-7) to gain access to the wires in the
connector. Measure the voltage (with the lamp turned on high) between the
orange and white wires.
Be careful, there should be voltage in excess of 100
Volts.
If the voltage is near 0: A cable or connector problem is indicated. Continue
with Step 4.
If the voltage is about 12V: The switching power supply (itÕs in the console)
for the lamp has failed. Contact LI-COR.
If the voltage is over 100V: (Normal)The problem may be a broken connec-
tion within the light source itself. Contact LI-COR.
4
Check the 26 pin D connector
Make sure none of the pins have been pushed in or broken (Table 20-1 on
page 20-45).
5
Check the cable
Try a different cable, if one is available to you.
Figure 20-7. Unscrew the lamp connector to access the four wires.
Caution -
High Voltage.
Содержание LI-6400
Страница 1: ...Using the LI 6400 Portable Photosynthesis System ...
Страница 15: ...Part I The Basics ...
Страница 16: ......
Страница 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Страница 175: ...Part II Useful Details ...
Страница 176: ......
Страница 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Страница 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Страница 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Страница 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Страница 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Страница 289: ...Part III Working With Files ...
Страница 290: ......
Страница 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Страница 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Страница 375: ...Part IV Configuration Issues ...
Страница 376: ......
Страница 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Страница 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Страница 455: ...Part V Maintenance Troubleshooting ...
Страница 456: ......
Страница 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Страница 587: ...Diagnostics and Utilities Sys Utility Programs Using the LI 6400 21 15 EBVpd t C guess EBSatVap _EBvap_kPa ...
Страница 593: ...Part VI Programming ...
Страница 594: ......
Страница 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Страница 846: ...Index I 16 Using the LI 6400 ...