Assembling the LI-6400
6400-01 CO2 Injector Installation
2-10
Using the LI-6400
2
6
Screw the cylinder cover onto the piercing block.
You may feel some resistance as the piercing pin contacts the cylinder. A
short burst of venting CO
2
may occur as the cylinder is pierced; the leak is
minimal if you continue to quickly tighten the cylinder cover. Tighten the
cover until snug; thereÕs no need to overtighten.
External CO
2
Tanks
The Tank Connector Block replaces the 6400-01 External CO
2
Source As-
sembly, and is useful in situations where CO
2
tanks or other high volume sup-
plies are available.
The tank connector block is designed for use at pressures between 180 and
220 PSIG (lbs in
-2
gauge pressure) of CO
2
.
Use a regulator, and do not ex-
ceed 250 PSIG CO
2
, as the pressure relief valve may vent your source
.
The Tank Connector Block is uses a 1/8Ó male NPT to 1/8Ó tubing fitting.
This fitting has a flow restrictor installed (10 cm
3
min
-1
).
Do not remove this
fitting
. A 1/8Ó to 4mm compression union is also provided for users who may
be unable to obtain 1/8Ó copper tubing. Directions for installing the Tank
Connector Block to a CO
2
source using 4mm copper tubing are given in
stallation Using 4mm Copper Tubing
■
■
■
■
To install the 9964-033 Assembly
1
Mount the CO
2
Tank Connector Block
Use the two knurled knobs to mound the block between the CO
2
and H
2
O
scrub tubes. Make sure that the O-ring seal on the back of the block is prop-
erly seated.
2
Insert copper tubing
Insert the 1/8Ó copper tubing between the 1/8Ó connector nut and the ferrule
(Figure 2-8).
Important
: Note the orientation of the ferrule. One of the tapered ends of the
ferrule is longer than the other; the long end must be oriented toward the con-
nector on the mounting block. When the nut is tightened onto the connector,
the ferrule will be permanently crimped to the copper tubing, and you will not
be able to remove it.
Содержание LI-6400
Страница 1: ...Using the LI 6400 Portable Photosynthesis System ...
Страница 15: ...Part I The Basics ...
Страница 16: ......
Страница 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Страница 175: ...Part II Useful Details ...
Страница 176: ......
Страница 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Страница 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Страница 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Страница 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Страница 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Страница 289: ...Part III Working With Files ...
Страница 290: ......
Страница 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Страница 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Страница 375: ...Part IV Configuration Issues ...
Страница 376: ......
Страница 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Страница 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Страница 455: ...Part V Maintenance Troubleshooting ...
Страница 456: ......
Страница 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Страница 587: ...Diagnostics and Utilities Sys Utility Programs Using the LI 6400 21 15 EBVpd t C guess EBSatVap _EBvap_kPa ...
Страница 593: ...Part VI Programming ...
Страница 594: ......
Страница 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Страница 846: ...Index I 16 Using the LI 6400 ...