Calibration Issues
View, Store Zeros & Spans
18-18
Using the LI-6400
18
The flow meterÕs zero has a slight temperature sensitivity, but the drift is typ-
ically less than 1 µmol s
-1
per 10 degrees C. The flow meter zero is controlled
by an analog output channel, and the value is displayed in the
"View, Store
Zeros & Spans"
program (page 18-18) under the Zeros column. The value
shouldnÕt change much from day to day, but checking it and setting it is so
easy, itÕs worth the effort to make sure.
View, Store Zeros & Spans
This entry in the Calib Menu launches a program that allows you to view the
current zero and span settings for the IRGAs, and the current flow meter zero.
You can also revert these values back to the power on or factory settings.
Store Reset Quit
Reset Options:
P - revert to PowerOn defaults
F - revert to Factory defaults
R - reset Reference IRGA
S - reset Sample IRGA
W - reset flow meter
Select a key
These values are the adjustment
terms
C
mr
and
C
ms
W
mr
and
W
ms
(14-6) on page 14-5. The values
should be between -2 and 2, typically.
These values are the gain factor
terms
G
cr
and
G
cs
G
wr
and
G
ws
in Equations (14-5) and (14-6) on
page 14-5. They should be between
0.95 and 1.05, typically.
These values are the
settings of the analog
output channels that
control offsets in the
IRGAs. These values
will be between
+/- 5000, but should
not be near either of
these extremes.
Figure 18-10. The ÒView,Store Zero&SpansÓ screen. The zero and span data for factory defaults and power on
defaults is stored in the file "/dev/parm1", along with a number of other settings.
Store the current
zero and span val-
ues as the Pow-
erOn Defaults
Set Flow Zero to 0.
Set sample IRGA zero to 0, span to 1.
Set reference IRGA zero to 0, span to 1.
PowerOn and Factory defaults are stored
in
“/dev/parm1”
.
µmol/mol
µmol/mol
mmol/mol
mmol/mol
Adjust
0.0
0.0
0.0
0.0
Calibration Parameters
1.012
1.007
0.987
0.991
Span
Zero
23.5
-23.5
42.5
107.9
220.1
CO2R:
CO2S:
H2OR:
H2OS:
FLOW:
Содержание LI-6400
Страница 1: ...Using the LI 6400 Portable Photosynthesis System ...
Страница 15: ...Part I The Basics ...
Страница 16: ......
Страница 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Страница 175: ...Part II Useful Details ...
Страница 176: ......
Страница 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Страница 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Страница 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Страница 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Страница 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Страница 289: ...Part III Working With Files ...
Страница 290: ......
Страница 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Страница 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Страница 375: ...Part IV Configuration Issues ...
Страница 376: ......
Страница 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Страница 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Страница 455: ...Part V Maintenance Troubleshooting ...
Страница 456: ......
Страница 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Страница 587: ...Diagnostics and Utilities Sys Utility Programs Using the LI 6400 21 15 EBVpd t C guess EBSatVap _EBvap_kPa ...
Страница 593: ...Part VI Programming ...
Страница 594: ......
Страница 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Страница 846: ...Index I 16 Using the LI 6400 ...