Primary Training
Calibration
3.3.5 End of calibration
Following the previously explained procedure, you can calibrate each channel of the
623 card. To calibrate the other 623 cards present in the BEN, first exit the program
using the [EXIT] command, wait until the acquisition error on the LCD screen of the
BEN disappears, then remove the RS-232 connector of the card and connect it to the
next card. Do not use the <escape> key to exit the program!
Warning
: NEVER remove the RS-232 cable before exiting the
program, otherwise calibration data will be lost.
After you disconnect the link between the PC and 623, the card goes on functioning
normally, and transmits the acquisition data to the processing unit.
If a channel is in the OFF position and you ask for a calibration process, the
following message appears on the screen:
The channel 5 is not enabled
RUN AutoConfig
OK