Calibration
Primary Training
TERMINALS
+
-
+
-
USER
SIGNAL
11A1
+
-
INPUT
CARD
11A11
USER SIDE
BEN SIDE
REFERENCE
VOLTAGE
SOURCE
4.1.2 Calibrate one voltage input without interruptible terminals (SSK110,
UK4,...)
1. Disconnect USER wires from both terminals
TERMINALS
+
-
+
-
USER
SIGNAL
11A1
+
-
INPUT
CARD
11A11
USER SIDE
BEN SIDE
REFERENCE
VOLTAGE
SOURCE
2. Short cut voltage inputs
Page 22