Calibration
Primary Training
When the range for each channel has been set up, quit the AUTOCONFIG window
by moving the arrows to the <OK> firld and press <ENTER>.
3.3
Calibrate menu
Place the cursor on the CALIBRATE menu, and select this choice with the
<ENTER> key.
3.3.1 Selection of the channels to be calibrated
A list representing the 8 channels appears on the screen:
SPACE to select AC/DC cal.
DC calibration
Channel 11A1 (Voltage )
Channel 11A2 (Voltage )
Channel 11A3 (Voltage )
Channel 11A4 (Voltage )
Channel 11A5 (Voltage )
Channel 11A6 (Voltage )
Channel 11A7 (Voltage )
Channel 11A8 (Voltage )
Calibration
Mode field
Select the calibration mode (DC mode, AC 50 Hz mode or AC 60 Hz mode) by
pressing the <SPACE> key while the calibration mode field is selected.
NOTE
:
When AC mode is selected, a sine-wave signal with a frequency of 50 Hz
±
0.1 Hz
or 60 Hz
±
0.1 Hz must be used.
The calibration can be carried out on one or up to four channels at a time, as long as
they are on the same range and have the same shunt value. Select the channels to be
calibrated with the vertical arrows and mark it with the space bar.