
23
For transmitters with rated maximum output power not listed above, the recommended separation distance d in metres (m)
may be determined using the equation adopted for the transmitter frequency, where P is the maximum rated output power
of the transmitter in Watts (W) stated by the transmitter manufacturer.
Notes:
(1) At 80 MHz and 800 MHz the highest frequency range applies.
(2) These guidelines might not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
Rated maximum
output power
of transmitter
(W)
from 150 kHz to 80 MHz
d = 1,2 √P
from 80 MHz to 800 MHz
d = 1,2 √P
from 800 MHz to 2,5 GHz
d = 2,3 √P
Separation distance (m) in relation to transmitter frequency
Warnings:
Although compliant with the EN 60601-1-2 standard, the
LTA420
medical device may
interfere with other
devices in the vicinity. The device must not be used in proximity to or stacked on top of
other equipment.
Install the device well away from other equipment that emits high frequencies (short waves,
microwaves,
electric scalpels, cell phones).
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are under
control. The customer or user can help prevent electromagnetic interference by maintaining a minimum distances
between mobile and portable RF communication equipment (transmitters) and the medical device as recommended
below, according to the maximum output power of the radio communication equipment.
Содержание Aspimed 1.9
Страница 37: ......