SYMBOLS USED
22
Type BF appliance
Switch On
Switch off (or battery on charge)
Direct current
Alternate current
Do not use the unit when takig a bath or a
shower
Conforms to Directive 93/42/EEC for
medical devices
The device contains a hermetic lead
battery. It must be disposed of in
accordance with current regulations
on the disposal of toxic-harmful
waste
Sterilisation by ethyl oxide
Disposable
It is compulsory to carefully read the instructions
before using this device
The device's casing is protected against
solids 12.5 mm or more in diameter, against
vertically falling drops of water and against
finger access to hazardous parts.
Electromagnetic Compatibility
Compliance levels according to EN 60601-1-2:2015 standard
- ESD immunity: 15 kV air, 8 kV contact (EN 61000-4-2)
- Burst immunity: 2 kV/100 kHz (EN 61000-4-4)
- Surge immunity (EN 61000-4-5): 1 kV common mode /2 kV differential mode
- Magnetic field (EN 61000-4-8): 30 A/m
- Immunity to rf currents in the 150 kHz-80 MHz range (EN 61000-4-6) 3 V modulation 80% 1 kHz
- RF emissions, CISPR 11: Class B
- Harmonics emissions, EN 61000-3-2: Class A
Rf field immunity (EN 61000-4-3):
Field (V/m)
Frequency
Modulation
3
80MHz 2700MHz
1kHz AM 80%
27
380MHz 390MHz
18Hz PM 50%
28
430MHz 470MHz
18Hz PM 50%
9
704MHz 787MHz
217Hz PM 50%
28
800MHz 960MHz
18Hz PM 50%
28
1700MHz 1990MHz
217Hz PM 50%
28
2400MHz 2570MHz
217Hz PM 50%
9
5100MHz 5800MHz
217Hz PM 50%
Содержание Aspimed 1.9