Axetris AG
LGD Compact Operating & Integration Instructions; Rev. D
41/46
Example 3 [50
ppm = 8
mA or 1.25
V; 100
ppm = 16
mA or 3.75
V]
min
max
delta
Concentration [ppm]
50
100
50
Analog voltage [V]
1.25
3.75
2.5
Analog current [mA]
8
16
8
normalized []
0.25
0.75
0.5
01
.
0
50
5
.
0
ppm
Span
25
.
0
25
.
0
01
.
0
50
Offset
For this example, the sensor parameters to type into the interface are:
AnalogOut_Min_Conc:
AnalogOut_Max_Conc:
AnalogOut_Offset:
AnalogOut_Span:
50
100
-0.25
0.01
Example 4 [50
ppm = 16
mA or 3.75
V; 100
ppm = 8
mA or 1.25
V]
min
max
delta
Concentration [ppm]
50
100
50
Analog voltage [V]
3.75
1.25
-2.5
Analog current [mA]
16
8
-8
normalized []
0.75
0.25
-0.5
01
.
0
50
5
.
0
ppm
Span
25
.
1
75
.
0
)
01
.
0
(
50
Offset
For this example, the sensor parameters to type into the interface are:
AnalogOut_Min_Conc:
AnalogOut_Max_Conc:
AnalogOut_Offset:
AnalogOut_Span:
50
100
1.25
-0.01