Axetris AG
LGD Compact Operating & Integration Instructions; Rev. D
25/46
Instructions for cable packaging
The LGD compact module will be delivered with an electrical cable from sensor housing to electronic boards
(all available cable types can be seen in chapter 1.2.2) The EMC conformity tests were done with the cable
packaging M12 straight D sub 200 mm shielded (Phoenix Contact 1430200).
Important: EMC conformity will only be granted for product executions with shielded cable versions together
with electronic box.
If this shielded cable (straight or angled version) cannot be used due to integration issues please be aware of
the following recommendations for the cable packaging:
•
Conductor cross section 17x 0.14 mm² (signal line), AWG 26
•
Fully shielded cable with good shield connection
•
Max. cable length = 350mm (in general as short as possible)
•
Twisted wire couples according to the following diagram:
Figure 19 Diagram of the cable wiring
Please be aware that the LGD performance is specified regarding our shielded cable versions and any
other cable packaging may create a loss in performance or an increasing noise level.
Figure 20 Pin connection of the LGD compact sensor module