Axetris AG
LGD Compact Operating & Integration Instructions; Rev. D
14/46
Sensor module
The LGD Compact-A module has two M4 threaded mounting holes located on the side (see Figure 2,
recommended to use two pan head cap screw M4x16 (1.5 Nm) for mounting). For gas supply there are
straight/angled QSML-M5-4 quick connectors for 4 mm diameter tubes.
Figure 2 LGD Compact-A module dimensions and interfaces for fixation and gas supply
Electronics
The electronics is located in a separate housing.
The outer dimensions of the electronic housing are 257 x 83 x 26mm (LxWxH) without connectors and screws.
To avoid damage during operation on the electronics it`s recommended to fix the electronics box with 4 pan
head cap screws M4x10 (1.5 Nm) at the ends.
The electronics housing contains two printed circuit boards (PCBs) as shown in Figure 3.
Figure 3 Electronics housing LGD Compact-A: interface board (left), main board (right)