29
Leica VT1000S – Vibrating-blade microtome
- The warning signal will cease
automatically after the initial-
ization phase.
- Retighten the loose clamping
screws.
If the clattering sound does
not cease once the clamping
screws have been retight-
ened, do not hesitate to call
the Technical Service imme-
diately.
Do not use the instrument
when in this condition.
- The visible clamping screws
have become loose during
sectioning.
These symptoms may occur
from time to time and are un-
avoidable, as the clamping
screws which have to be op-
erated by the user cannot be
sealed.
6.
Malfunctions: Meanings and Troubleshooting
Error messages/symptoms
Sources of error
Troubleshooting
- Audible warning signal.
(When operating the instru-
ment for the first time or after
the E-EPROM has been ex-
changed.)
- A clattering sound can be
heard.