Leica VT1000S – Vibrating-blade microtome
- Lower the specimen disc suffi-
ciently to prevent collision.
- Lower the specimen disc suffi-
ciently to prevent collision.
When working with orient-
able specimen discs, move
the buffer tray to its lowest
position directly after switch-
ing on the instrument!
Unlock the
REV/FORW
-button
by pulling it back to the center
position.
- To reactivate the return stroke
movement, lock the
REV/FORW
-button again
(to
REV
position).
- Unlock the
REV/FORW
-button
by pulling it back to the center
position.
- To reactivate the return stroke
movement, lock the
REV/FORW
-button again
(to REV position).
6.
Malfunctions: Meanings and Troubleshooting
Error messages/symptoms
Sources of error
Troubleshooting
- Collision of knife and specimen
holder.
- Audible warning signal.
- Return stroke is not completed.
- Audible warning signal.
- Return stroke is not completed.
- Clearance angle adjustment:
- If a clearance angle wider than
5° is selected, specimen disc
and knife edge can potentially
collide with each other.
- When working with orientable
specimen holders, knife edge
and specimen holder can col-
lide at any selected clearance
angle.
----- Operating error due to locking
function of the REV/FORW
button:
- With the
REV/FORW
-button
locked the instrument is
switched off via the main
switch at the rear of the instru-
ment and is switched on again
without releasing the
REV/FORW
-button to its center
position.
- With the
REV/FORW
-locked the
instrument was switched off
via the emergency Stop and
after that the emergency stop
was released again without
releasing the
REV/FORW
-but-
ton to its center position.