20
Operating Manual V1.8 – 12/2009
Adjusting the retraction
•
In programming mode, press
DISP
to display the specimen retraction
menu.
•
The indication 'LO' will be displayed.
•
Press the
-/+
button to set a specimen retraction value between 1 and
999 µm. Or to turn off the retraction by selecting '0'.
•
The selected value will be displayed in the FEED window.
•
Press CLR to quit the menu function.
5.4
Adjusting the amplitude
•
To obtain excellent sectioning results, the amplitude requires adjustment
according to the specimen type being sectioned.
To this end:
•
With a 2.5 mm Allen key loosen the clamping screw
(1)
and secure the
eccentric on the bottom with your finger.
Selectable amplitude positions are, from left to right:
0.2 mm; 0.4 mm; 0.6 mm; 0.8 mm; 1 mm.
•
Slide the amplitude clamping screw to the desired amplitude position
and retighten.
When adjusting the amplitude setting, do not remove the clamping
screw, simply loosen it.
The instrument is shipped with the amplitude set to 0.6 mm.
5.
Operation
Fig. 20
1