26
Operating Manual V1.8 – 12/2009
Error messages/symptoms
Sources of error
Troubleshooting
- Audible warning signal.
- Downward stroke is not
completed.
- Audible warning signal.
- Downward stroke is not
completed.
- The sectioning motor stops.
- Any processing step (section-
ing stroke etc.) is interrupted
immediately.
- Any UP/DOWN motion of the
buffer tray is interrupted imme-
diately.
- Any locked buttons are indicat-
ed by an audible sound signal.
- When pressing any key, the
instrument gives an audible
sound signal.
- In case the
Emergency stop
function
has been activated,
the instrument will remain inop-
erational when pressing the
foot switch.
- The indication "
SP
" is displayed.
- Operating error due to locking
function of the UP/DOWN-
button:
- With the
UP/DOWN
-button
locked in the DOWN position
the instrument was switched
off via the main switch at the
rear of the instrument switched
on again without releasing the
UP/DOWN
-button to its center
position
- With the
UP/DOWN
button
locked the instrument was
switched off via the emergency
stop (foot switch or Emergency
stop button) and after that the
emergency stop was released
without unlocking the
UP/
DOWN
-button.
- The
Emergency stop function
has been activated.
- Release the
UP/DOWN
-button
to its center position.
- To reactivate the downward
motion, activate the
UP/DOWN
-
button again (
DOWN
).
- Release the
UP/DOWN
-button
to its center position.
- To reactivate the downward
motion, activate the
UP/DOWN
-
button again (
DOWN
).
- Release the
Emergency stop
button
.
- Select an operating mode and
continue working.
6.
Malfunctions: Meanings and Troubleshooting
Содержание VT1000S
Страница 2: ......