Operating Manual Leica EM ACE600
Version 01/19
Page 89
Firmly tighten the screws, but do not overtighten.
If the screws to clamp the rods are not tightened enough the rod can slip
through, the connection between tip and flat part will not be under spring
force and evaporation is not possible.
When taking out a rod to exchange and/or sharpen it, open the screw (1) and
push down. The clamp will open and the rod can be easily taken out.
Use the special tool to keep the carbon rod evaporation source (spring loaded)
open.