Operating Manual Leica EM ACE600
Version 01/19
Page 45
4.3
Vacuum test
After connecting the coater as described in 3.5, the coater can be tested for
vacuum.
Switch on the instrument on the rear by pushing the mains switch down and push
the Pump button.
If the vacuum does not reach 100
mbar in 5 min, a vacuum leak is present
and needs to be eliminated.
4.4
Process start and stop
The coating process can be started at any time (system evacuated or not) if the
door and source cover are closed and the source is connected (electronic safety
switches prohibit starting with open door or source cover).
The
button will turn into a
button when the system starts the
coating cycle.
Once the “Start” button is activated, the coater will automatically run the complete
coating cycle. At the end of the process the system eithers stays under vacuum or
vents automatically (vent after process is activated, defined within the process
parameters).
Pressing the “Stop” button will terminate the process (after confirmation)
regardless of the step of the process. The system either stays under vacuum or
vents automatically (vent after process is activated, defined within the process
parameters)
4.5
Thickness monitoring and geometrical correction (QSG)
A quartz crystal swings in a certain frequency. Coated with a material, this
frequency reduces according to the material and the applied thickness. With this
information, the accurate film thickness coated during a process run can be
determined.
Every Leica EM ACE600 system is equipped with the QSG film thickness
measurement monitor. The user can choose between termination by QSG and