Operating Manual Leica EM ACE600
Version 01/19
Page 50
If not all screws are tightened the instrument may not recognize the thread
section.
Clean the sealing surface with a lint-free tissue.
There is a Leica EM ACE600 YouTube video where the loading of the thread
can be seen: http://www.youtube.com/watch?v=Qj3Y-WfNbvM
After the carbon thread is loaded, replace the head, gently tighten the fastening
screws and connect the cables and close the cover.
Do not tighten the screws when the instrument is under vacuum. They can’t
be opened when vented and there is a risk of damaging the thread.