134
Hazards of use
WARNING
The absence of instruction, or the inadequate
imparting of instruction, can lead to incorrect or
adverse use, and can give rise to accidents with far-
reaching human, material, financial and environ-
mental consequences.
Precautions:
All users must follow the safety directions given by
the manufacturer and the directions of the person
responsible for the instrument.
WARNING
Using a battery charger not recommended by
Leica Geosystems can destroy the batteries. This
can cause fire or explosions.
Precautions:
Only use chargers recommended by Leica Geosy-
stems to charge the batteries.
CAUTION
Watch out for erroneous measurement
results if the product has been dropped or has been
misused, modified, stored for long periods or trans-
ported.
Precautions:
Periodically carry out test measurements and per-
form the field adjustments indicated in the User
Manual particularly after the instrument has been
subjected to abnormal use and before and after
important measurements.
CAUTION
Strong magnetic fields in the immediate vici-
nity (e.g. transformers, melting furnaces...) may
influence the compensator and lead to measuring
errors.
Counter measures:
When measuring near strong magnetic fields, check
results for plausibility.