5. OPER ATION - STAGES AND OBJECT DISPL ACEMENT
5.6.2 Adjusting the step sizes on motorized stages
t 0O UIF 5PVDI 4DSFFO VTF UIF
key to switch to
stage and focus controls. The X/Y coordinates of the cur-
rent stage position are displayed.
t 4FMFDUUIF
Stage
menu using the
key.
t:PVDBODIBOHF
UIF
travel speed of the stage by
switching between the
Coarse
and
Fine
step sizes.
If the
Coarse
value is selected, the travel speed is the
same for all objectives.
The
Fine
value is adapted to the respective objective.
5.6.3
Saving and moving to stage positions
In the LAS X software, the programmed working area is
marked by a dashed line frame.
You can reset the working area to the
maximum stage
travel range via the
Limits
key.
Working area restriction
Stored
stage position
Current
stage position
Zero point
1
3
2
52
Содержание DMi8