5. OPER ATION - TR ANSMIT TED LIGHT CONTR AST ME THOD
Manual integrated fluorescence axis
1 Manual filter change
2 Analyzer
3 Manual shutter
4 Manual setting for FIM
5.14 Fluorescence
Note:
To replace filter cubes, refer to Chapter 5.11, Filters
o
. S.
63
Leica DMi8 with manual fluorescence
t *G OFDFTTBSZ mSTU TXJUDI PWFS UP UIF
fluorescence axis
(
Fluor
) by opening the shutter (pull out switch rod (3) for
manual shutter or push shutter button for motorized shut-
ter).
t *GOFDFTTBSZQVMMPVUUIFBOBMZ[FSTMJEFS DPNQMFUFMZUP
open the beam path.
t 4FUUIFMJHIUJOUFOTJUZPOUIF
Leica EL6000 compact light
source.
Note:
If the Leica DMi8 is equipped with a 2 port infinity port (
o
S. 44)
switch over to the current port .
o
S. 58
t 8IFOVTJOHUIF
Fluorescence Intensity Manager (FIM):
The brightness is adjusted in 5 defined increments on the
FIM slider (4).
100% / 55% / 30% / 17% / 10%
t 5IF
filter cubes are swiveled manually into the beam path
by turning the fluorescent turret disk (1).
t "DPMPSNBSLPOUIFFEHFPGUIFEJTLJOEJDBUFTUIFDVS
-
rent filter cube. If the shipped configuration is changed by
inserting another filter cube, note the color of the position
where the filter cube was inserted so you can allocate it
later.
Manual external fluorescence axis
3 Manual shutter
4 Manual setting for FIM
5 Connection for light guide
6 Diaphragm module
7 Receptacle for EFW or structured illumination
5
4
6
3
7
1
2
10
4
3
82