5. OPER ATION - TR ANSMIT TED LIGHT CONTR AST ME THOD
Toggling
Transmitted Light/fluorescence
Shutter control
Note:
If all positions of the fluorescent turret disk have been as-
signed, the "DAPI" filter cube can be replaced with the "DA-
PI-TL" filter cube using LAS X software. Transmitted Light
contrast methods are possible with this filter cube.
5.13.1 Brightfield (TL)
Leica DMi8 with manual Transmitted Light method
t *GOFDFTTBSZTFUUIF5-#SJHIUmFMEQPTJUJPOPSBOFNQUZ
position on the fluo turret, or select the "DAPI-TL" filter
cube.
For a tiltable Transmitted Light arm:
t4FMFDUUIF#SJHIUmFMEQPTJUJPOPOUIFDPOEFOTFS
BF
.
For a fixed Transmitted Light arm:
t 'PS DPSSFDU IFJHIU BEKVTUNFOU PG UIF 4 BOE
S40/0.45 condensers, the stand has marks (1). These
marks pertain to a liquid height of 15 mm. For stands
with double marks, the upper and lower mark repre-
sent a range with different liquid heights. Move the
condenser holder while observing the specimen to at-
tain an optimal image.
Note:
For the stands with 3-plate mechanical stages, posi-
tion the arm 25 mm below the marks, as the illumina-
tion axis is installed 25 mm higher on an adapter than
on the fixed stage.
t 1SFTT UIF TUPQ MFWFS BOE BEKVTU UIF 5SBOTNJUUFE
Light illumination arm until the upper edge of the car-
rier is even with the corresponding condenser height
mark.
t 4XJOHBMMSFNBJOJOHPQUJDBMDPNQPOFOUTTVDIBTUIFBOB
-
lyzer, polarizer or IC prisms, from the beam path.
t*OTFSUB5SBOTNJUUFE-JHIUTQFDJNFO
t4FMFDUZPVSPCKFDUJWF
67